| ITEM |
SPECIFICATIONS |
| Titration method |
Volumetric Karl Fischer reagent titration method |
| Titration control method |
Intermittent addition of titrant while approaching the end point.
(1) Control using a waiting interval timer of 1 - 99sec (INTERVAL).
Titration stops when there is no change in the electrode potential during this interval.
(2) Control by titration timer. (T-TIMER: 1 - 99min.)
|
| End point detection |
DC polarization potential difference detection. |
| Buret capacity |
10ml
reproducibility ±0.01 ml (for full discharge, standard deviation for n=5) |
| Liquid contact material |
PTFE, ceramic and glass |
| Number of burets |
1 |
| Display screen |
2 lines by 32digits with backlit LCD |
| Display units |
ml, mg H2O, %, or ppm (display minimum=0.01ml) |
| Measuring range |
100 ppm - 100% |
| Minimum volume of addition |
0.01ml |
| Calculation function |
Concentration, factor standardization |
| Titration cell capacity |
150ml |
| Minimum volume of solvent |
Approx. 30ml |
| External input/output |
Two RS-232C ports
Port 1 (HOST) for PC connection.
Port2 (BALANCE) for Balance connection. |
| Printers |
Optional thermal and impact printer - please inquire. |
| Power supply ; Power consumption |
AC100 -240 V, 50/60 Hz & 100 VA |
| Dimensions |
W310mm x D270mm x H300mm |
| Weight |
About 6kg |